An AI stethoscope missed most heart murmurs in cats, an NC State study finds

In a 51-cat exam, the human-trained device flagged 2 of 22 murmurs. The August paper says the result should not be generalized to every veterinary AI tool.

DailyMech Staff

An AI stethoscope missed most heart murmurs in cats, an NC State study finds
AI-generated illustration (Grok Imagine) for DailyMech — Original AI-generated image

An AI stethoscope built for people did a poor job of finding heart murmurs in cats, according to a study from North Carolina State University published online August 5, 2026, in the Journal of the American Veterinary Medical Association. The university's news release is dated August 12. The counts below come from that release and from dvm360's account of the paper. The paper itself, DOI 10.2460/javma.26.05.0353, was not opened for this article.

Researchers examined 105 animals, 54 dogs and 51 cats. A board-certified veterinary cardiologist, a cardiology resident, and a fourth-year student each listened. The cardiologist's exam was the reference. The device's AI was trained on human data, the NC State team and dvm360 said.

In dogs, clinicians heard murmurs in 38 of 54 animals, about 70 percent. The device matched 33 of those 38, about 87 percent. It correctly flagged all six dogs with atrial fibrillation and also labeled 22 other dogs as having it. The study found the device never classified a dog as free of an arrhythmia, so it had no reliable normal category for rhythm.

Cats were worse. Clinicians heard murmurs in 22 of 51 cats, about 43 percent. The device detected two of those and missed 20. Fourth-year students identified nearly two-thirds of the murmurs in the affected cats. The first author described the concerning result as a tool used for reassurance that failed to find a real murmur.

The authors did not call the device useless. They said it may still make useful recordings, and that automated interpretations need caution in veterinary patients. They said the findings should not be extended to every digital or AI stethoscope. The limits are plain: a small sample, one device, and a clinician's ear rather than a confirmed diagnosis as the standard.

The adoption context is moving faster than this kind of test. A Digitail and American Animal Hospital Association benchmark released October 1, 2026, surveying 1,730 professionals in the United States and Canada, said 83.7 percent use AI, up from 39.2 percent in 2024. That is self-reported use, not accuracy. David Brundage's March 5, 2026 audit in Frontiers in Veterinary Science noted that veterinary AI tools do not face the federal premarket review that a similar human device often would. DailyMech covered a different accuracy question, abdominal X-rays rather than heart sounds, in the JAVMA pilot, and the school giveaway that prompted some of that scrutiny in the Vetology article.

A result in the other direction, so the species-and-task point stays visible: a Texas A&M study in PLOS One, described in a university release the same day as the NC State release, found an AI tool measuring heart size on dog chest X-rays about as accurately as trained veterinarians. Owner attitudes are a third topic. The November HABRI and Chewy survey is already reported in the pet-owner piece.

This is a study write-up, not veterinary advice.